Others

Spectrophotometer Cells / Cuvettes

Scientific Instruments has a large range of Spectrophotometer cells available.

The most common type used is the Type 1 G 10 this is the standard type cell in glass.

Material Code Wavelength Usual spectral range of application
Optical Glass G 334 to 2300nm Visible light
Infrasil Quartz I 220 to 3800nm Near Infrared
Spectrosil Quartz Q 170to 2600nm UV region

If you are measuring below 350nm then a Quartz cell is required for the light to

transmit through the specimen. I.e. Type 1 Q 10.

Matched sets of cells can be ordered normally 2 or 4.

We can also supply Demountable Cells & Micro Flow Cells for your specific type of work.

Please download this Spectrophotometer cells / cuvettes specifications PDF and contact us for delivery and pricing.

ORDER / ENQUIRE MORE INFO

^ Back to top

UV1100

Optics Seya-Namioka mount monochromator, Ratio beam principle
Wavelength range 190nm~1100nm
Spectral bandwidth 4 nm (UV 1100)
1 nm (UV 1101)
2 nm (UV 1102)
Stray light ≤0.05% T (220nm NaI, 340nm NaNO2)
Wavelength scan speed 10, 100, 200, 400, 800, 1200, 2400, 3600 nm/min
Wavelength accuracy ≤ ±0.5nm
Wavelength setting repeatability ≤ ±0.5nm (UV 1100)
≤ ±0.3nm (UV 1101, UV1102)
Wavelength slew 6000 nm/min
Baseline Stability 0.001Abs/hr (500nm, 2 hours later)
Photometric range Abs: – 3 to 3
%T: 0 to 600
Conc: 0.000 to 9999
Photometric accuracy ±0.002 Abs (0 to 0.5 Abs)
±0.004 Abs (0.5 to 1 Abs)
≤±0.3 %T
Photometric repeatability ±0.001 Abs (0 to 0.5 Abs)
±0.002 Abs (0.5 to 1 Abs)
≤±0.15 %T
Noise level 0.0005Abs (500nm) (UV 1100)
0.0008Abs (500nm) (UV 1101, UV1102)
Light Source WI and D2 lamps
Light source switching Automatic switching at 340nm
Detector Silicon photodiode
Display LCD with back-light (94mmx70mm)
Printer Parallel port, supported printing language of PCL3 and ESC/P
Dimension (L x W x H)mm 370 x 550 x 265
Weight (kg) 20
Power requirement 100 to 240 switchable, 50/ 60 Hz

ORDER / ENQUIRE MORE INFO

^ Back to top